Exploration of oxygen-mediated disinfection of medical units reveals a excessive sensitivity of Pseudomonas aeruginosa to elevated oxygen ranges


Micro organism and progress circumstances

For the current proof-of-principle research, we utilized well-characterized and available bacterial kind strains to facilitate inter-laboratory comparisons. Specifically, the next strains had been used all through the experiments: Escherichia coli ATCC 25,922, Staphylococcus aureus HG-001, Enterococcus faecalis ATCC 29,212, Enterococcus faecalis ATCC 51,299, Klebsiella pneumoniae ATCC 11,228 and P. aeruginosa ATCC 27,853. Every pressure was saved in 20% glycerol and frozen at -80 °C. Ranging from the frozen shares, every pressure was plated on blood agar (BA) plates and grown in a single day at 37 °C. The next day, 5–6 colonies had been chosen to inoculate 20 ml of lysogeny broth (LB; Oxoid). 100 ml glass bottles had been used to tradition the micro organism at 37 °C with 250 rpm shaking for 4 h. Acceptable dilutions had been carried out to acquire a closing beginning inoculum with a measured optical density at 600 nm (OD600) of 0.05. Of observe, we utilized OD600 measurements as a normal for the bacterial cell density all through the current research, as an alternative of the choice McFarland normal22,23.

Bacterial oxygen susceptibility assay

The totally different micro organism had been grown as described above. A 1 × phosphate-buffered saline (PBS) tradition suspension, equivalent to 2–4 × 106 colony-forming models (CFU)/mL was chosen as beginning inoculum for the assay. Utilizing 24-well plates, triplicates consisting of 20 µL of inoculum had been unfold throughout the underside of particular person wells. The plate was then allowed to air-dry, with the lid off for 1.5 h in a laminar air circulation (LAF) cupboard. After drying, the plate was transferred to a custom-built gasoline incubator (Fig. 1). The incubator lid was closed on the chamber, and subsequently flushed for five min with the chosen oxygen combination at 0.8 kPa. Subsequently, the gasoline circulation was stopped, and each the inlet and outlet of the incubator had been closed. Remedy occasions of 1, 5, 10, or 30 min at room temperature (RT) had been used. After the therapy, the incubators had been opened, and the 24-well plates had been incubated at RT for 20 h. Every properly was then washed with 600 µL of PBS 1x, to gather the micro organism. Lastly, the remaining viable micro organism had been quantified by means of serial dilutions, plating on BA, and colony counting, which allowed calculation of the respective CFU/mL.

Determine 1
figure 1

Experimental setup of gasoline incubation chambers. Ranging from (1), the gasoline cylinder is linked to a strain valve (2). A plastic tube (3) connects the strain valve to the incubator inlet (4). On the incubator, additionally a strain valve is current (5). Inside the incubator (6) the experimental agar plate (7) or 24-well microtiter plate (not proven) is positioned. As soon as the incubator is sealed with the lid and flushed with the suitable gasoline combination, the gasoline circulation is allowed by means of an outlet (8) situated on the polar reverse of the inlet. Lastly, an adjustable tube is linked to the outlet that may be closed (9).

Gasoline incubators and oxygen mixtures

Gasoline incubators as proven in Fig. 1 had been manufactured by the instrument makers of the College Medical Heart Groningen (UMCG; ‘Analysis Instrumenten-makerij’). The incubators had been designed in an oblong form, with dimensions of 300 mm × 140 mm, and a top of 40 mm. Every incubator was designed to have an inlet and an outlet to permit the circulation of gasses, and a security strain valve. The incubators had been composed of two separate components, the chamber and the lid. The chambers had been sealed with the lids by means of 4 metallic clamps situated on the facet of every chamber. The oxygen concentrations chosen for the experiments within the current research had been: 42%, 53%, 63% and 87%, the oxygen being solely complemented with nitrogen. As well as, management experiments had been performed with common air, in what follows known as 21% oxygen.

Nebulizers

The Ventobra nebulizer24, commercialized by PARI Pharma and provided by Westfalen A.G., was chosen as a consultant medical gadget for the experimental oxygen-mediated disinfection course of. This gadget is ‘semi-critical’ in line with the Spaulding standards. The Ventobra system was not too long ago accepted by the European Medicines Company to be used by CF sufferers (EMA/169,512/2015 Web page 3/3). In our research, particular consideration was given to the Tolero® handset for the Ventobra nebulizer, used as a goal for various disinfection protocols.

Microbial contamination and decontamination of nebulizers

The Tolero handset was disassembled into its three distinct components, particularly a plastic membrane (MB), a mouth-piece (MP) and a metallic piece (MT), as proven in Fig. 2. The bacterium chosen for the contamination experiment was P. aeruginosa ATCC 27,853, grown as described above. The surfaces of the three separated gadget components had been purposely contaminated by bringing them into contact for ~ 10 s with a bacterial suspension of two–4 × 106 CFU/mL in PBS (OD600 of 0.05). The three contaminated gadget components the place then allowed to air-dry in a LAF cupboard and subsequently positioned in gasoline incubators for therapy with a steady flux of the very best oxygen focus examined, 87% O2, for 30 min (0.8 kPa). After the therapy, two approaches had been adopted to evaluate the numbers of micro organism on the surfaces of the gadget components. The ‘stamping’ method entailed the contact software of every nebulizer half onto a BA plate. In distinction, the ‘swabbing’ method entailed the gathering of micro organism from the gadget components’ surfaces with a sterile cotton swab and subsequent streaking of BA plates. The BA plates had been then incubated in a single day at 37 °C and bacterial progress was inspected on the subsequent day.

Determine 2
figure 2

The handset of the Tolero nebulizer. The higher a part of the picture exhibits the totally assembled handset, whereas the decrease a part of the picture exhibits its three important elements upon disassembly, particularly the mouthpiece (MP), membrane (MB), and the metallic half (MT).

For management, each the stamping and swabbing strategies had been utilized in two totally different approaches. Firstly, the three purposely contaminated gadget components had been disinfected by means of immersion in both 70% ethanol or 30% hydrogen peroxide for 10 min, air-dried in a LAF cupboard for 10 min and examined for bacterial contamination. This allowed us to confirm that the gadget components had been free from micro organism previous to their re-use. Secondly, to make sure that no traces of disinfectant had been current on the gadget surfaces prior bacterial (re-)contamination, which may confound the outcomes of our experiments, the components that had been disinfected with 70% ethanol or 30% hydrogen peroxide had been immersed in sterile MilliQ water for 10 s.

Statistical evaluation

Outcomes are introduced in imply + / − normal deviation. The statistical evaluation of the info was carried out with GraphPad Prism 8.0.1 (GraphPad Software program, USA), the place a p-value < 0.05 was thought-about important. The take a look at utilized was two-way Anova.